Agilent 1200 DAD
Detector type1024-element diode array
Light sourceDeuterium and tungsten lamp
Short-term noise< ± 0.8 x 10 -5 AU at 254 nm and at 750 nm
Drift< 0.9 x 10 -3 AU/hr, at 254 nm
Linearity>2 AU upper limit
Wavelength range190-950 nm
Wavelength accuracy± 1 nm, self-calibration with deuterium lines, verification with holmium oxide filter
Slit widthProgrammable: 1, 2, 4, 8, 16 nm
Diode width< 1 nm
Wavelength bunchingProgrammable, 2 - 400 nm, in steps of 1 nm
Flow cells
Standard13-µL volume, 10 mm cell path length, 120 bar (1760 psi) pressure maximum
Semi-micro5-µL volume, 6 mm cell path length, 120 bar (1760 psi) pressure maximum
Micro2-µL volume, 3 mm cell path length, 400 bar (5880 psi) pressure maximum
Semi-nano500-nanoliter volume 10-mm cell path length, 50 bar (735 psi) pressure maximum
Nano80 nanoliter volume., 6-mm cell path length, 50 bar (735 psi) pressure maximum
Preparative3-mm cell path length, 120 bar (1760 psi) pressur maximum0.3-mm cell path lenght, 20 bar (294 psi) pressure maximum, quartz0.06-mm cell path length, 20 bar (294 psi) pressure maximum, quartz
High pressure1.7-µL volume, 6-mm cell path length, 400 bar (5880 psi) pressure maximum
Time programmableWavelength, polarity, peak width, lamp bandwidth, autobalance,WL range, threshold, spectra storage mode
Spectral toolsData analysis software for spectra evaluation, including spectral libraries and peak purity functions
Analog outputRecorder/integrator 100 mV or 1 V, 2 outputs
GLPData recovery card to prevent data losses. RFID for electronics records of flow cell and UV lamp conditions (path length, volume, product number, serial number, test passed, usage). Electronic records of maintenance and errors. Early maintenance feedback (EMF) for continuous tracking of instrument usage, display of feedback messages if preset limits are exceeded. Verification of wavelength accuracy with built-in holmium oxide filter.
Preis auf Anfrage
Agilent 1200 DAD SL
Dieser gebrauchte Agilent DAD SL der Serie 1200 (G1315C) wird generalüberholt geliefert.
Detector type1024-element diode array
Light sourceDeuterium and tungsten lamp
Short-term noise< ± 0.8 x 10 -5 AU at 254 nm and at 750 nm
Drift< 0.9 x 10 -3 AU/hr, at 254 nm
Linearity>2 AU upper limit
Wavelength range190-950 nm
Wavelength accuracy± 1 nm, self-calibration with deuterium lines, verification with holmium oxide filter
Slit widthProgrammable: 1, 2, 4, 8, 16 nm
Diode width< 1 nm
Wavelength bunchingProgrammable, 2 - 400 nm, in steps of 1 nm
Flow cells
Standard13-µL volume, 10 mm cell path length, 120 bar (1760 psi) pressure maximum
Semi-micro5-µL volume, 6 mm cell path length, 120 bar (1760 psi) pressure maximum
Micro2-µL volume, 3 mm cell path length, 400 bar (5880 psi) pressure maximum
Semi-nano500-nanoliter volume 10-mm cell path length, 50 bar (735 psi) pressure maximum
Nano80 nanoliter volume., 6-mm cell path length, 50 bar (735 psi) pressure maximum
Preparative3-mm cell path length, 120 bar (1760 psi) pressur maximum0.3-mm cell path lenght, 20 bar (294 psi) pressure maximum, quartz0.06-mm cell path length, 20 bar (294 psi) pressure maximum, quartz
High pressure1.7-µL volume, 6-mm cell path length, 400 bar (5880 psi) pressure maximum
Time programmableWavelength, polarity, peak width, lamp bandwidth, autobalance,WL range, threshold, spectra storage mode
Spectral toolsData analysis software for spectra evaluation, including spectral libraries and peak purity functions
Analog outputRecorder/integrator 100 mV or 1 V, 2 outputs
GLPData recovery card to prevent data losses. RFID for electronics records of flow cell and UV lamp conditions(path length, volume, product number, serial number, test passed, usage).Electronic records of maintenance and errors. Early maintenance feedback (EMF) for continuous tracking of instrument usage,display of feedback messages if preset limits are exceeded. Verification of wavelength accuracy with built-in holmium oxide filter.
Preis auf Anfrage
ESA Corona Detector + Column Compartment
• Reliable and simple to operate.• Universal detection of nonvolatile and many semi-volatile analytes.• Four orders of magnitude dynamic range (nonlinear).• Low ng lower limits of detection.• Typically < 2% R.S.D. response variability.• Response magnitude is minimally dependent on the analyte’s chemical properties.• Broad applicability.• Isocratic and gradient compatibility.
Preis auf Anfrage
ESA Coul Array Heated Organizer 70-4470T
Temperature controlled enclosure for maintaining constant temperature and organizing coulometric array cells, pre-filters, column(s), manual injector valve, PEEK pulse dampers and biocompatible gradient mixers.Temperature range: + 5°C above ambient to 45°CTemperature accuracy: <, 0.5°C
Preis auf Anfrage
ESA Coulochem II
MODELCOULOCHEM II HPLC SYSTEM
DETECTOR (SCATTERED LIGHT)
MODELUV/VIS MODEL 522
Wavelengths, nm (OPERATING)190-800programmable
Drift, absorb/hr (DETECTOR)<,0.0002
Noise, aufs (DETECTOR)<,±0.00001
Flow cell vol, µl (DETECTOR)0.25-15
Cell pathlength, mm (DETECTOR)2-10, varies with
Avail absorb, aufs (DETECTOR)0.0005-3
Recorder output (DETECTOR)Selectable
Event marker (DETECTOR)Yes
SYSTEM CONFIGURATION
Design (SYSTEM CONFIGURATION)Modular, isocratic
std configurationDepends on config
SAMPLE INJECTORAutosampler - for details please see comments.542 Autosampler
Fixed loop or
variable (Fixed loop or)Variable
Manual or automatic (SYSTEM CONFIGURATION)Automatic
SAMPLE VOLUME, µL1
Amount of vials (SYSTEM CONFIGURATION)24,84,96
Injections/sample (SYSTEM CONFIGURATION)9
Perfomance time, min (SYSTEM CONFIGURATION)9hr 59 min 59 sec
Maximal pressure (SYSTEM CONFIGURATION)5000
Auto flushing (SYSTEM CONFIGURATION)Yes
Standard vial (SYSTEM CONFIGURATION)4types
Kind of vial (SYSTEM CONFIGURATION)0.5,1.5, 4,10mL
DATA MANAGEMENT
MODELELITE EZCHROM FOR
Channels (POWER NEEDED)4
DISPLAYMS Windows-based
Chromatograms,
no. stored (Chromatograms,)Unlimited
Optional storage (DATA MANAGEMENT)Yes
PUMPESA-582
Type (CONFIGURATION)Dual piston,
Pump heads (PUMP)2
Flow range, ml/min (ELUENT PUMP)Up to 10 - please see comments for detailed specifications.0.001-9.999
Flow increm, ml/min (PUMP)0.001
Flow precision, % (ELUENT PUMP)0.3
Flow accuracy, % (ELUENT PUMP)1
Max pressure, psi (PUMP)5000
Max/min pressure
switches (Max/min pressure)Yes
Gradient capability (PUMP)Optional with second
Solvents (PUMP)1isocratic
OTHER DETECTORS (DATA MANAGEMENT)Coulochem Electro-
COLUMN (SPECIFICATIONS)
MODELMETA-250 : MICRODI-
Construction (ROTOR)Stainless steel
Length/internal
Diameter, mm (UMBILICAL CORD)250:150:80:150
Particle size, µm (SPECIFICATIONS)37320
Particle type (COLUMN)Silica, polymeric
Sorbent type (SPECIFICATIONS)
Reverse phase (COLUMN)C18, C8, proprietary
OTHER SPECIFICATIONS (AIR SUPPLY)Manual injector.
CE MARK (MDD) (Interference compensation)Yes
Preis auf Anfrage
ESA Coulochem II
MODELCOULOCHEM II HPLC SYSTEM
DETECTOR (SCATTERED LIGHT)
MODELUV/VIS MODEL 522
Wavelengths, nm (OPERATING)190-800programmable
Drift, absorb/hr (DETECTOR)<0.0002
Noise, aufs (DETECTOR)<±0.00001
Flow cell vol, µl (DETECTOR)0.25-15
Cell pathlength, mm (DETECTOR)2-10, varies with
Avail absorb, aufs (DETECTOR)0.0005-3
Recorder output (DETECTOR)Selectable
Event marker (DETECTOR)Yes
SYSTEM CONFIGURATION
Design (SYSTEM CONFIGURATION)Modular, isocratic
std configurationDepends on config
SAMPLE INJECTORAutosampler - for details please see comments.542 Autosampler
Fixed loop or
variable (Fixed loop or)Variable
Manual or automatic (SYSTEM CONFIGURATION)Automatic
SAMPLE VOLUME, µL1
Amount of vials (SYSTEM CONFIGURATION)24,84,96
Injections/sample (SYSTEM CONFIGURATION)9
Perfomance time, min (SYSTEM CONFIGURATION)9hr 59 min 59 sec
Maximal pressure (SYSTEM CONFIGURATION)5000
Auto flushing (SYSTEM CONFIGURATION)Yes
Standard vial (SYSTEM CONFIGURATION)4types
Kind of vial (SYSTEM CONFIGURATION)0.5,1.5, 4,10mL
DATA MANAGEMENT
MODELELITE EZCHROM FOR
Channels (POWER NEEDED)4
DISPLAYMS Windows-based
Chromatograms,
no. stored (Chromatograms,)Unlimited
Optional storage (DATA MANAGEMENT)Yes
PUMPESA-582
Type (CONFIGURATION)Dual piston,
Pump heads (PUMP)2
Flow range, ml/min (ELUENT PUMP)Up to 10 - please see comments for detailed specifications.0.001-9.999
Flow increm, ml/min (PUMP)0.001
Flow precision, % (ELUENT PUMP)0.3
Flow accuracy, % (ELUENT PUMP)1
Max pressure, psi (PUMP)5000
Max/min pressure
switches (Max/min pressure)Yes
Gradient capability (PUMP)Optional with second
Solvents (PUMP)1isocratic
OTHER DETECTORS (DATA MANAGEMENT)Coulochem Electro-
COLUMN (SPECIFICATIONS)
MODELMETA-250 : MICRODI-
Construction (ROTOR)Stainless steel
Length/internal
Diameter, mm (UMBILICAL CORD)250:150:80:150
Particle size, µm (SPECIFICATIONS)37320
Particle type (COLUMN)Silica, polymeric
Sorbent type (SPECIFICATIONS)
Reverse phase (COLUMN)C18, C8, proprietary
OTHER SPECIFICATIONS (AIR SUPPLY)Manual injector.
CE MARK (MDD) (Interference compensation)Yes
Preis auf Anfrage
ESA Coulochem II
MODELCOULOCHEM II HPLC SYSTEM
DETECTOR (SCATTERED LIGHT)
MODELUV/VIS MODEL 522
Wavelengths, nm (OPERATING)190-800programmable
Drift, absorb/hr (DETECTOR)<0.0002
Noise, aufs (DETECTOR)<±0.00001
Flow cell vol, µl (DETECTOR)0.25-15
Cell pathlength, mm (DETECTOR)2-10, varies with
Avail absorb, aufs (DETECTOR)0.0005-3
Recorder output (DETECTOR)Selectable
Event marker (DETECTOR)Yes
SYSTEM CONFIGURATION
Design (SYSTEM CONFIGURATION)Modular, isocratic
std configurationDepends on config
SAMPLE INJECTORAutosampler - for details please see comments.542 Autosampler
Fixed loop or
variable (Fixed loop or)Variable
Manual or automatic (SYSTEM CONFIGURATION)Automatic
SAMPLE VOLUME, µL1
Amount of vials (SYSTEM CONFIGURATION)24,84,96
Injections/sample (SYSTEM CONFIGURATION)9
Perfomance time, min (SYSTEM CONFIGURATION)9hr 59 min 59 sec
Maximal pressure (SYSTEM CONFIGURATION)5000
Auto flushing (SYSTEM CONFIGURATION)Yes
Standard vial (SYSTEM CONFIGURATION)4types
Kind of vial (SYSTEM CONFIGURATION)0.5,1.5, 4,10mL
DATA MANAGEMENT
MODELELITE EZCHROM FOR
Channels (POWER NEEDED)4
DISPLAYMS Windows-based
Chromatograms,
no. stored (Chromatograms,)Unlimited
Optional storage (DATA MANAGEMENT)Yes
PUMPESA-582
Type (CONFIGURATION)Dual piston,
Pump heads (PUMP)2
Flow range, ml/min (ELUENT PUMP)Up to 10 - please see comments for detailed specifications.0.001-9.999
Flow increm, ml/min (PUMP)0.001
Flow precision, % (ELUENT PUMP)0.3
Flow accuracy, % (ELUENT PUMP)1
Max pressure, psi (PUMP)5000
Max/min pressure
switches (Max/min pressure)Yes
Gradient capability (PUMP)Optional with second
Solvents (PUMP)1isocratic
OTHER DETECTORS (DATA MANAGEMENT)Coulochem Electro-
COLUMN (SPECIFICATIONS)
MODELMETA-250 : MICRODI-
Construction (ROTOR)Stainless steel
Length/internal
Diameter, mm (UMBILICAL CORD)250:150:80:150
Particle size, µm (SPECIFICATIONS)37320
Particle type (COLUMN)Silica, polymeric
Sorbent type (SPECIFICATIONS)
Reverse phase (COLUMN)C18, C8, proprietary
OTHER SPECIFICATIONS (AIR SUPPLY)Manual injector.
CE MARK (MDD) (Interference compensation)Yes
Preis auf Anfrage
ESA Coulochem II Model 5200A
MODELCOULOCHEM II HPLC SYSTEM
DETECTOR (SCATTERED LIGHT)
MODELUV/VIS MODEL 522
Wavelengths, nm (OPERATING)190-800programmable
Drift, absorb/hr (DETECTOR)<0.0002
Noise, aufs (DETECTOR)<±0.00001
Flow cell vol, µl (DETECTOR)0.25-15
Cell pathlength, mm (DETECTOR)2-10, varies with
Avail absorb, aufs (DETECTOR)0.0005-3
Recorder output (DETECTOR)Selectable
Event marker (DETECTOR)Yes
SYSTEM CONFIGURATION
Design (SYSTEM CONFIGURATION)Modular, isocratic
std configurationDepends on config
SAMPLE INJECTORAutosampler - for details please see comments.542 Autosampler
Fixed loop or
variable (Fixed loop or)Variable
Manual or automatic (SYSTEM CONFIGURATION)Automatic
SAMPLE VOLUME, µL1
Amount of vials (SYSTEM CONFIGURATION)24,84,96
Injections/sample (SYSTEM CONFIGURATION)9
Perfomance time, min (SYSTEM CONFIGURATION)9hr 59 min 59 sec
Maximal pressure (SYSTEM CONFIGURATION)5000
Auto flushing (SYSTEM CONFIGURATION)Yes
Standard vial (SYSTEM CONFIGURATION)4types
Kind of vial (SYSTEM CONFIGURATION)0.5,1.5, 4,10mL
DATA MANAGEMENT
MODELELITE EZCHROM FOR
Channels (POWER NEEDED)4
DISPLAYMS Windows-based
Chromatograms,
no. stored (Chromatograms,)Unlimited
Optional storage (DATA MANAGEMENT)Yes
PUMPESA-582
Type (CONFIGURATION)Dual piston,
Pump heads (PUMP)2
Flow range, ml/min (ELUENT PUMP)Up to 10 - please see comments for detailed specifications.0.001-9.999
Flow increm, ml/min (PUMP)0.001
Flow precision, % (ELUENT PUMP)0.3
Flow accuracy, % (ELUENT PUMP)1
Max pressure, psi (PUMP)5000
Max/min pressure
switches (Max/min pressure)Yes
Gradient capability (PUMP)Optional with second
Solvents (PUMP)1isocratic
OTHER DETECTORS (DATA MANAGEMENT)Coulochem Electro-
COLUMN (SPECIFICATIONS)
MODELMETA-250 : MICRODI-
Construction (ROTOR)Stainless steel
Length/internal
Diameter, mm (UMBILICAL CORD)250:150:80:150
Particle size, µm (SPECIFICATIONS)37320
Particle type (COLUMN)Silica, polymeric
Sorbent type (SPECIFICATIONS)
Reverse phase (COLUMN)C18, C8, proprietary
OTHER SPECIFICATIONS (AIR SUPPLY)Manual injector.
CE MARK (MDD) (Interference compensation)Yes
Preis auf Anfrage
Waters Acquity UPLC FLR Detector
Wellenlängenbereich200 bis 890 nm (Anregung) 210 bis 900 nm (Emission)
Wellenlängengenauigkeit±3 nm
Bandbreite20 nm
Wellenlängenreproduzierbarkeit±0,25 nm
EmpfindlichkeitSignal/Rausch-Verhältnis > 1000 (Raman-Spektrum von H2O)
Messbereich0,001 bis 10.000 EU
DatenaufnahmeBis zu 80 Hz
Aufbau der FlusszelleAxial beleuchtet
Zellvolumen< 2 μL
Druckgrenzwert500 psi
Materialien mit FlüssigkeitskontaktQuarzglas, FEP, PEEK
Abmessungen und GewichtBreite: 34,3 cm (13,5 Zoll)Höhe: 21,6 cm (8,5 Zoll)Tiefe: 50,8 cm (20,0 Zoll)
Preis auf Anfrage
Waters Acquity UPLC FLR Detector
Wellenlängenbereich200 bis 890 nm (Anregung) 210 bis 900 nm (Emission)
Wellenlängengenauigkeit±3 nm
Bandbreite20 nm
Wellenlängenreproduzierbarkeit±0,25 nm
EmpfindlichkeitSignal/Rausch-Verhältnis > 1000 (Raman-Spektrum von H2O)
Messbereich0,001 bis 10.000 EU
DatenaufnahmeBis zu 80 Hz
Aufbau der FlusszelleAxial beleuchtet
Zellvolumen< 2 μL
Druckgrenzwert500 psi
Materialien mit FlüssigkeitskontaktQuarzglas, FEP, PEEK
Abmessungen und GewichtBreite: 34,3 cm (13,5 Zoll)Höhe: 21,6 cm (8,5 Zoll)Tiefe: 50,8 cm (20,0 Zoll)
Preis auf Anfrage
Waters TQ Detector
Zwei hochauflösende Quadrupol-Analysatoren (MS1/MS2) sowie Vorfilter zur Maximierung von Auflösung und Transmission bei gleichzeitiger Vermeidung von Kontamination der Hauptanalysatoren.T-Wave™1 aktiviert für optimale MS/MS-Leistung bei hohen Datenerfassungsraten.Rauscharmer, achsenunabhängiger Photomultiplier-Detektor mit langer LebensdauerDigitaler Dynamikbereich bis zu 4 x 10^6.Voller MS ScanProdukt-Ionen-ScanVorläufer-Ionen-ScanKonstanter NeutralitätsverlustAusgewählte Ionenaufzeichnung (SIR)Mehrfach-Reaktionsüberwachung (MRM)Umfrage-Scan-Modi: Vollscan MS-getriggerter ProduktionenscanMassenbereich 2 bis 2048 m/zScan-Geschwindigkeit Bis zu 10.000 Da/sBeispiele für erreichbare Erfassungsraten 10 Scans pro Sekunde (m/z 100 bis 1000), 20 Scans pro Sekunde (m/z 50 bis 500)Massenstabilität Die Massendrift beträgt <0,1 Da über einen Zeitraum von 24 Stunden
Preis auf Anfrage
Young Lin / Schambeck Acme 9000 RI750F (Refractive Index Detector)
RI Range1.00 to 1.75 RIU
Typical flow rate0.2 ~ 3.0 ml/min
Noise≤ 5 x 10¹⁰ RIU
Auto zero range40 x 10⁵ RIU
Auto zero resolutionto 5 x 10¹⁰ RIU
Recorder output± 1 V
Linear dynamic range80 x 10⁵ RIU
Drift rate4 x 10⁸ RIU
Flow rate range0.2 ~ 3.0 mL/min
Cell volume9 μl
Cell pressure6 kg/cm² (84 psi)
Volume into cell24 μl
Temperature control35°C ~ 55°C
Digital interfaceRS-232C
Digital inputPurge, Auto Zero
Dimension297 × 198 × 495 mm (WxHxD)
Power requirement110/220 ± 10% VAC. 50-60 Hz.
Weight15 kg