Produktinformationen "Flow Heated Column Compartment Columns: 1"

For better separation in preparative Liquid Chromatography

Each parameter in the analytical LC is set and monitored
very precisely. The temperature of the column is an
essential factor for a good separation and is adjusted
depending on the problem and analytes. In the preparative
field, temperature control for columns has been thought of,
but columns are only heated externally. For low flow rates
in the analytical field this might be sufficient, but for
preparative separations and high flow rates a lot of
efficiency is lost when the mobile phase is not heated.
  • Lower mobile phase viscosity
  • Faster interaction with stationary phase
  • Lower system back pressure
  • Higher flowrates without loosing resolution
  • Shorter cycle times
  • Equalize the enthalpy of mixing solvents
 
ColumnsUp to 8 columns (8x 30x250mm or 2x 50x250mm)ControlIntegrated Keyboard control, PC direct control
Column selection Valve8-Pos. Max Pressure: 5000psiOutputsError signal contact closure
Flowrate200ml/minInputsBCD-/Impulse Signal for Column position
Max. Pressure400bar1/16" SS-Tubing Heat ExchangerDimensions (WxHxD)35x61x28 cm
Temperature Range5 - 55 °C (in 0.5 °C increments)Weight27 kg
Accuracy0.1 °CEnviromental Conditions5 - 40 °C, 35 - 75% rH
ReproducibilityBetter than 0.5 °CPower Requirements115/230 VAC; 50/60 Hz
StabilityBetter than 0.5 °CMax. power Consumption1520 VA
SensorsSolvent Vapor alarm, Leakage alarm, Temperature warning