Flow Heated Column Compartment
Columns:
1
For better separation in preparative Liquid ChromatographyEach parameter in the analytical LC is set and monitoredvery precisely. The temperature of the column is anessential factor for a good separation and is adjusteddepending on the problem and analytes. In the preparativefield, temperature control for columns has been thought of,but columns are only heated externally. For low flow ratesin the analytical field this might be sufficient, but forpreparative separations and high flow rates a lot ofefficiency is lost when the mobile phase is not heated.Lower mobile phase viscosityFaster interaction with stationary phaseLower system back pressureHigher flowrates without loosing resolutionShorter cycle timesEqualize the enthalpy of mixing solvents
ColumnsUp to 8 columns (8x 30x250mm or 2x 50x250mm)ControlIntegrated Keyboard control, PC direct control
Column selection Valve8-Pos. Max Pressure: 5000psiOutputsError signal contact closure
Flowrate200ml/minInputsBCD-/Impulse Signal for Column position
Max. Pressure400bar1/16" SS-Tubing Heat ExchangerDimensions (WxHxD)35x61x28 cm
Temperature Range5 - 55 °C (in 0.5 °C increments)Weight27 kg
Accuracy0.1 °CEnviromental Conditions5 - 40 °C, 35 - 75% rH
ReproducibilityBetter than 0.5 °CPower Requirements115/230 VAC; 50/60 Hz
StabilityBetter than 0.5 °CMax. power Consumption1520 VA
SensorsSolvent Vapor alarm, Leakage alarm, Temperature warning