For better separation in preparative Liquid ChromatographyEach parameter in the analytical LC is set and monitoredvery precisely. The temperature of the column is anessential factor for a good separation and is adjusteddepending on the problem and analytes. In the preparativefield, temperature control for columns has been thought of,but columns are only heated externally. For low flow ratesin the analytical field this might be sufficient, but forpreparative separations and high flow rates a lot ofefficiency is lost when the mobile phase is not heated.Lower mobile phase viscosityFaster interaction with stationary phaseLower system back pressureHigher flowrates without loosing resolutionShorter cycle timesEqualize the enthalpy of mixing solvents
Columns1 columnup to 50x250mm)ControlIntegrated Keyboard control, PC direct control
Column selection Valve8-Pos. Max Pressure: 5000psiOutputsError signal contact closure
Flowrate200ml/minInputsBCD-/Impulse Signal for Column position
Max. Pressure400bar1/16" SS-Tubing Heat ExchangerDimensions (WxHxD)35x61x28 cm
Temperature Range5 - 55 °C (in 0.5 °C increments)Weight27 kg
Accuracy0.1 °CEnviromental Conditions5 - 40 °C, 35 - 75% rH
ReproducibilityBetter than 0.5 °CPower Requirements115/230 VAC; 50/60 Hz
StabilityBetter than 0.5 °CMax. power Consumption1520 VA
SensorsSolvent Vapor alarm, Leakage alarm, Temperature warning
The valve shown in the picture is not included in the scope of delivery and must be ordered separately from the list shown below.
Product NumberFL-V200FL-V201FL-V204FL-V512FL-V239FL-V05
Type6 port 2 position3 port 2 way2 port selection6 port 2 position
Max Pressure20MPa (2900psi)30MPa (4350psi)1MPa (145psi)35 Mpa (5800 psi)
Wetted MaterialsPEEKSUS, PEEKPEEKSUS, PEEK
Tube OD1/16"ø3, 1/8"1/16"
Path Diameter0.6mm (0.024")0.8mm (0.03")0.8mm (0.03")1.0mm (0.04")0.6 mm
Thread SizeNo.10-32UNF1/4-28UNFNo.10-32UNF
Solvent Degassing SystemA trouble-less degassing series equipped with larger internal volume and real time displayFeaturesPick the model with the right internal volume for your applicationAuto cleaning function prevents vacuum pump deteriorationCPU control achieves stable degassingIndependent chamber design prevents solvent contaminationChoose from 2 channel or 4 channel configurationsReal time display of vacuum level
Model number
FL-AG-42-01
FL-AG-44-01
FL-BG-42-02
FL-BG-44-02
FL-BG-42-10
FL-BG-44-10
FL-FG-42
FL-PG-41P
FL-PG-42P
Channels
2
4
2
4
2
4
2
1
2
Internal volume
340 µl
510 µl
1020 µl
10 ml
62.0 ml
Flow rate
≤ 10 ml/min
≦ 10 ml/min
≦ 5ml/min
≦ 50.0 ml/min
Vacuum range
100 hPa fixed
150 hPa fixed
Thread size
1/4-28 UNF
Wetted material
Teflon TM AF*1, ETFE, PTFE
PTFE, ETFE, PPS
PTFE, ETFE, PEEK
Remote control
-
Input: - Output: Error signal
Dimensions (excluding protuberances)
108 (H) × 70 (B) × 240 (T)
mm
180(H) × 90(W) × 400(D) mm
Weight
approx. 2.1 kg
approx. 2.3 kg
approx. 5.0 kg
approx. 6.0 kg
Power
AC 100 - 240 V (50 / 60 Hz) MAX 80 VA
*1:Teflon™ AF is a registered trademark of the Chemours Corporation.Note: cannot be used with fluorinated solvents.
The valve shown in the picture is not included and must be purchased separately from the list below.It cannot be attached to FLOM pumps.
Product NumberFL-V200FL-V201FL-V204FL-V512FL-V239FL-V05
Type6 port 2 position3 port 2 way2 port selection6 port 2 position
Max Pressure20MPa (2900psi)30MPa (4350psi)1MPa (145psi)35 Mpa (5800 psi)
Wetted MaterialsPEEKSUS, PEEKPEEKSUS, PEEK
Tube OD1/16"ø3, 1/8"1/16"
Path Diameter0.6mm (0.024")0.8mm (0.03")0.8mm (0.03")1.0mm (0.04")0.6 mm
Thread SizeNo.10-32UNF1/4-28UNFNo.10-32UNF
Intelligent Micro Pump MP-22Valve Switching Micro PumpCheck valve-less design for super low pulse control under low flow ratesFeaturesImproved near-zero pulse flowProblems with bubble greatly reducedEasy setupUpper and lower pressure limitersSpace-saving compact designExternal control available with RS232C protocol
Part numberMP-22-110S
Pumphead materialSUS
Max pressure*35 MPa (5075 psi)
Flow rate range**0.001 - 9.999 mL/min
Flow per stroke80 µL
Plunger stroke10 mm
Flow accuracy***< 0.3 %
FeaturesActive pulse reduction, upper/lower pressure limiters
Pressure accuracy< 5 %
Wetted materialsSapphire, Ruby, PTFE, PCTFE, PFA, ETFE, FFKM, HDPE, PEEK, SUS
Remote controlInput:PUMP ON / OFF, Output:Error signal
CommunicationRS232C, RS485(between FLOM devices)
Dimensions150 (H) ×, 160 (W) ×, 280 (D) mm, excluding protuberances
WeightApprox. 6.7 kg
PowerAC100 - 240 V (50 / 60 Hz) MAX 140 VA
*: Maximum pressure is only sustained intermittently.**: Maximum output is not sustainable over long periods of time.***: Using Methanol at flow rate10~500&mu,L/min , output pressure >, 1 MPa ~ 20MPa
This intelligent pump of the UI-22 series is now being replaced by the intelligent pump of the UI-32 series. However, spare parts for the UI-22 series are still available.FeaturesPulses reduced to absolute minimum making low flow rate pumping possibleBubble problems greatly reducedLong life plunger sealUpper and lower pressure limitersSpace-saving compact designControl with RS232C protocolChoose SUS, PEEK, or PCTFE to fit your solvent
Part numberUI-22-110SUI-22-110PUI-22-110DUI-22-410SUI-22-410PUI-22-410D
Pumphead materialSUSPEEKPCTFESUSPEEKPCTFE
Max pressure*35 MPa (5075 psi)20 MPa (2900 psi)2 MPa (290 psi)5 MPa (725 psi)5 MPa (725 psi)2 MPa (290 psi)
Flow rate range**0.001 - 9.999 mL/min0.01 - 99.99 mL/min
Flow per stroke80 μL708 μL
Plunger stroke10 mm
Flow accuracy<, 0.3 %
FeaturesActive pulse reduction, upper/lower pressure limiters
Pressure accuracy<, 5 %
Wetted materialsSUS: Sapphire, Ruby, PTFE, PCTFE, PFA, FFKM, HDPE, PEEK, SUS
PEEK: Sapphire, Ruby, PTFE, PCTFE, PFA, ETFE, FFKM, HDPE, PEEK
PCTFE: Sapphire, Ruby, PTFE, PCTFE, PFA, ETFE, FFKM
Remote controlInput:PUMP ON / OFF, Output:Error signal
CommunicationRS232C , RS485(between FLOM devices)
Dimensions144 (H) × 105 (W) × 240 (D) mm, excluding protuberances
WeightApprox. 5.6 kg
PowerAC100 - 240 V(50 / 60 Hz) MAX140 VA
*: Maximum pressure is only sustained intermittently.**: Maximum output is not sustainable over long periods of time.
A convenient, ultra-compact valve for switching flow pathsFeatures• Attach to the valve panel (sold separately) to allow operation from valve head side • A line-up of 5 units to cover a wide variety of applications
Product NumberFL-V200FL-V201FL-V204FL-V512FL-V239FL-V05
Type6 port 2 position3 port 2 way2 port selection6 port 2 position
Max Pressure20MPa (2900psi)30MPa (4350psi)1MPa (145psi)35 Mpa (5800 psi)
Wetted MaterialsPEEKSUS, PEEKPEEKSUS, PEEK
Tube OD1/16"ø3, 1/8"1/16"
Path Diameter0.6mm (0.024")0.8mm (0.03")0.8mm (0.03")1.0mm (0.04")0.6 mm
Thread SizeNo.10-32UNF1/4-28UNFNo.10-32UNF
Corresponding side panels for the valves can be ordered under the item numbers shown in the list below:
Product NameSide Panel
Product NumberFL-4512FL-4513FL-4514
Pump TypeUI Series, MP Series, SP Series, DP-11
Attachable PartsVI-11, VI-12, MI-11,V05V200, V201, V204, V239, V512
Damper Attachableyesnoyes
Dimensions144(H)×60(W)×44(D), excluding protuberances
A convenient, ultra-compact valve for switching flow pathsFeatures• Attach to the valve panel (sold separately) to allow operation from valve head side • A line-up of 5 units to cover a wide variety of applications
Product NumberFL-V200FL-V201FL-V204FL-V512FL-V239FL-V05
Type6 port 2 position3 port 2 way2 port selection6 port 2 position
Max Pressure20MPa (2900psi)30MPa (4350psi)1MPa (145psi)35 Mpa (5800 psi)
Wetted MaterialsPEEKSUS, PEEKPEEKSUS, PEEK
Tube OD1/16"ø3, 1/8"1/16"
Path Diameter0.6mm (0.024")0.8mm (0.03")0.8mm (0.03")1.0mm (0.04")0.6 mm
Thread SizeNo.10-32UNF1/4-28UNFNo.10-32UNF
Corresponding side panels for the valves can be ordered under the item numbers shown in the list below:
Product NameSide Panel
Product NumberFL-4512FL-4513FL-4514
Pump TypeUI Series, MP Series, SP Series, DP-11
Attachable PartsVI-11, VI-12, MI-11,V05V200, V201, V204, V239, V512
Damper Attachableyesnoyes
Dimensions144(H)×60(W)×44(D), excluding protuberances
A convenient, ultra-compact valve for switching flow pathsFeatures• Attach to the valve panel (sold separately) to allow operation from valve head side • A line-up of 5 units to cover a wide variety of applications
Product NumberFL-V200FL-V201FL-V204FL-V512FL-V239FL-V05
Type6 port 2 position3 port 2 way2 port selection6 port 2 position
Max Pressure20MPa (2900psi)30MPa (4350psi)1MPa (145psi)35 Mpa (5800 psi)
Wetted MaterialsPEEKSUS, PEEKPEEKSUS, PEEK
Tube OD1/16"ø3, 1/8"1/16"
Path Diameter0.6mm (0.024")0.8mm (0.03")0.8mm (0.03")1.0mm (0.04")0.6 mm
Thread SizeNo.10-32UNF1/4-28UNFNo.10-32UNF
Corresponding side panels for the valves can be ordered under the item numbers shown in the list below:
Product NameSide Panel
Product NumberFL-4512FL-4513FL-4514
Pump TypeUI Series, MP Series, SP Series, DP-11
Attachable PartsVI-11, VI-12, MI-11,V05V200, V201, V204, V239, V512
Damper Attachableyesnoyes
Dimensions144(H)×60(W)×44(D), excluding protuberances
A convenient, ultra-compact valve for switching flow pathsFeatures• Attach to the valve panel (sold separately) to allow operation from valve head side • A line-up of 5 units to cover a wide variety of applications
Product NumberFL-V200FL-V201FL-V204FL-V512FL-V239FL-V05
Type6 port 2 position3 port 2 way2 port selection6 port 2 position
Max Pressure20MPa (2900psi)30MPa (4350psi)1MPa (145psi)35 Mpa (5800 psi)
Wetted MaterialsPEEKSUS, PEEKPEEKSUS, PEEK
Tube OD1/16"ø3, 1/8"1/16"
Path Diameter0.6mm (0.024")0.8mm (0.03")0.8mm (0.03")1.0mm (0.04")0.6 mm
Thread SizeNo.10-32UNF1/4-28UNFNo.10-32UNF
Corresponding side panels for the valves can be ordered under the item numbers shown in the list below:
Product NameSide Panel
Product NumberFL-4512FL-4513FL-4514
Pump TypeUI Series, MP Series, SP Series, DP-11
Attachable PartsVI-11, VI-12, MI-11,V05V200, V201, V204, V239, V512
Damper Attachableyesnoyes
Dimensions144(H)×60(W)×44(D), excluding protuberances
A convenient, ultra-compact valve for switching flow pathsFeatures• Attach to the valve panel (sold separately) to allow operation from valve head side • A line-up of 5 units to cover a wide variety of applications
Product NumberFL-V200FL-V201FL-V204FL-V512FL-V239FL-V05
Type6 port 2 position3 port 2 way2 port selection6 port 2 position
Max Pressure20MPa (2900psi)30MPa (4350psi)1MPa (145psi)35 Mpa (5800 psi)
Wetted MaterialsPEEKSUS, PEEKPEEKSUS, PEEK
Tube OD1/16"ø3, 1/8"1/16"
Path Diameter0.6mm (0.024")0.8mm (0.03")0.8mm (0.03")1.0mm (0.04")0.6 mm
Thread SizeNo.10-32UNF1/4-28UNFNo.10-32UNF
Corresponding side panels for the valves can be ordered under the item numbers shown in the list below:
Product NameSide Panel
Product NumberFL-4512FL-4513FL-4514
Pump TypeUI Series, MP Series, SP Series, DP-11
Attachable PartsVI-11, VI-12, MI-11,V05V200, V201, V204, V239, V512
Damper Attachableyesnoyes
Dimensions144(H)×60(W)×44(D), excluding protuberances
A convenient manual valve for switching between flow paths.FeaturesLow torque design for easy switching
Product NumberFL-V200FL-V201FL-V204FL-V512FL-V239FL-V05
Type6 port 2 position3 port 2 way2 port selection6 port 2 position
Max Pressure20MPa (2900psi)30MPa (4350psi)1MPa (145psi)35 Mpa (5800 psi)
Wetted MaterialsPEEKSUS, PEEKPEEKSUS, PEEK
Tube OD1/16"ø3, 1/8"1/16"
Path Diameter0.6mm (0.024")0.8mm (0.03")0.8mm (0.03")1.0mm (0.04")0.6 mm
Thread SizeNo.10-32UNF1/4-28UNFNo.10-32UNF
Great for plunger pump cleaning and basic pumping applicationsFeatures:Space-saving compact designTubes can be selected according to the solvent usedTube cassette can be set in four directions
Product NameRoller Pump
ModelWP-21
Flow rate range4-stage variable (approx. 0.5・1.0・2.0・3.0mL/min)
Flow Rate Switching 4-stage via front switch (rotary type)
Special Function Purge delivery via long press of front switch (approx. 5mL/min)
Flow Accuracy± 5 % (PharMed®BPT*¹ when new)
Discharge Pressure 90 kPa or less (PharMed®BPT*¹)
Suction Pressure-90 kPa or less (PharMed®BPT*¹)
Operating Liquid Temperature Range 5 to 50°C
Rotation DirectionClockwise when viewed from front
External Input/Output Signals Input: Conductive OFF, Open ON Output: Error output via open collector
Tube MaterialPharMed®BPT*¹ / FLURAN F-5500-A / Transmaster TM-15
Dimensions 65 (H) × 46 (W) × 110 (D) mm, excluding protrusions
WeightApprox. 0.5 kg
Power DC 24 V, MAX 15 W (AC adapter included)
*1 “PharMed®BPT” is a registered trademark of Saint-Gobain K.K.* Standard specifications use “PharMed®BPT” tubing. Please inquire if you require “Fulan F-5500-A” or "TransMasterTM-15”. * Turning the power switch ON automatically initiates fluid delivery.